Fast DAQ system with image rejection for axion dark matter searches

S. Ahn, M. J. Lee, A. K. Yi, B. Yeo, B. R. Ko, Y. K. Semertzidis

Preprint: not peer reviewed. Results may change in later versions.

Abstract

A fast data acquisition (DAQ) system for axion dark matter searches utilizing a microwave resonant cavity, also known as axion haloscope searches, has been developed with a two-channel digitizer that can sample 16-bit amplitudes at rates up to 180 MSamples/s. First, we realized a practical DAQ efficiency of greater than 99% for a single DAQ channel, where the DAQ process includes the online fast Fourier transforms (FFTs). Using an IQ mixer and two parallel DAQ channels, we then also implemented a software-based image rejection without losing the DAQ efficiency. This work extends our continuing effort to improve the figure of merit in axion haloscope searches, the scanning rate.

Abstract: arXiv descriptive metadata (CC0). Thank you to arXiv for use of its open access interoperability. This service was not reviewed or approved by, nor does it necessarily express or reflect the policies or opinions of, arXiv.

Record

Archive ID
arxiv:2109.08312
Published
Sep 17, 2021
Updated
Apr 18, 2022
DOI
10.1088/1748-0221/17/05/p05025
arXiv
2109.08312
Categories
physics.ins-det, hep-ex
Provenance
arxiv