Abstract
Brain-Computer Interfaces (BCIs) face a severe calibration bottleneck due to cross-subject spatial covariance shifts and physiological artifacts. To enable zero-calibration BCI, a deep learning pipeline was engineered combining Per-Session Independent Component Analysis, Riemannian Euclidean Alignment, and EEGNet stabilized by Stochastic Weight Averaging (SWA). Evaluated on the strict MOABB BNCI2014-001 benchmark, the proposed architecture successfully isolates true sensorimotor rhythms. For the primary case study (Subject 1), a clinically robust SWA stable accuracy of 90.97% (AUC: 0.976, Cohen's $κ$: 0.819) was achieved. Furthermore, expanded 9-fold Leave-One-Subject-Out (LOSO) cross-validation yielded a globally stable mean accuracy of 74.31%, proving hardware-agnostic zero-shot efficacy for binary motor imagery.
Abstract: arXiv descriptive metadata (CC0). Thank you to arXiv for use of its open access interoperability. This service was not reviewed or approved by, nor does it necessarily express or reflect the policies or opinions of, arXiv.
Record
- Archive ID
- arxiv:2607.16225
- Published
- Jun 19, 2026
- arXiv
- 2607.16225
- Categories
- eess.SP, cs.LG, q-bio.NC
- Provenance
- arxiv