A Novel Magnetoscan Setup

S. Haindl

Preprint: not peer reviewed. Results may change in later versions.

Abstract

Due to a modification in the original magnetoscan setup, a significant improvement in resolution was obtained. The paper focuses on experimental results which should support the idea of the new setup using two magnets with opposite direction of magnetization. This contribution to the characterization techniques of melt-grown bulk superconductors should promote the easy installation of this technique in industry. The improved magnetoscan technique may further help to investigate growth-induced inhomogeneities of the top-seeded-melt-growth process in the submillimeter range, and it offers new possibilities to the characterization of smaller structures such as superconducting films or coated conductors.

Abstract: arXiv descriptive metadata (CC0). Thank you to arXiv for use of its open access interoperability. This service was not reviewed or approved by, nor does it necessarily express or reflect the policies or opinions of, arXiv.

Record

Archive ID
arxiv:cond-mat/0508416
Published
Aug 18, 2005
Venue
Supercond. Sci. Technol. 18 (2005) 1483-1488
DOI
10.1088/0953-2048/18/11/013
arXiv
cond-mat/0508416
Categories
cond-mat.supr-con
Provenance
arxiv